15
Measurement Operation (3)
C. Measuring Resistance (See figure 5)
Warning
To avoid damages to the Meter or to the devices under test,
disconnect circuit power and discharge all the high-voltage
capacitors before measuring resistance.
The resistance ranges are: 400.0Ω , 4.000kΩ , 40.00kΩ , 400.0kΩ ,
4.000MΩ and 40.00MΩ . To measure resistance, connect the Meter
as follows:
1. Insert the red test lead into the
HzΩ terminal and the black
test lead into the COM terminal.
2. Model UT205: Set the rotary switch to Ω.
Model UT206: Set the rotary switch to
Ω , resistance
measurement (Ω) is default or press BLUE button to select Ω
measurement mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
• The test leads can add 0.1Ω to 0.2Ω of error to resistance
measurement. To obtain precision readings in low-resistance
measurement, that is the range of 400.0Ω , short-circuit the input
terminals beforehand, using the relative measurement function button
REL
∆
to automatically subtract the value measured when the
testing leads are short-circuited from the reading.
• If Ω reading with shorted test leads is not ≤0.5Ω , check for
loose test leads, incorrect function selection, or enabled Data Hold
function.
• For high-resistance measurement (>1MΩ ), it is normal to take
several seconds to obtain a stable reading.
• The LCD displays
indicating open-circuit for the tested resistor
or the resistor value is higher than the maximum range of the Meter.
red
black
red
black
Model UT205/UT206: OPERATING MANUAL