Accuracy Specification(2)
Remarks:
l At 10A Range:
For continuous measurement 10 seconds and
interval time between 2 measurement greater than
15 minutes.
Remarks:
l Frequency response: 40Hz – 400Hz.
l At 10A Range:
For continuous measurement 10 seconds and
interval time between 2 measurement greater than
15 minutes.
l Displays effective value of sine wave (mean value
response).
C. DC Current
200µA
2mA
20mA
200mA
10A
0.1µA
1µA
10µA
0.1mA
10mA
Range Resolution
(0.8%+2)
Overload
Protection
Accuracy
(1.2%+5)
Fuse 315mA,
250V, fast type,
5x20mm
Fuse 10A,
250V, fast type,
5x20mm
D. AC Current
200µA
2mA
20mA
200mA
10A
0.1µA
1µA
10µA
0.1mA
10mA
Range Resolution
(1.0%+5)
Overload
Protection
Accuracy
(2.0%+5)
Fuse 315mA,
250V, fast type,
5x20mm
Fuse 10A,
250V, fast type,
5x20mm