4. Calibration
page 45
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
4.5.3 Cal. Concentration
F2- Automatic Calibration is selected in the Calibration Mode menu.
Now press F1-Cal. Concentration to enter concentrations used for calibration. Up to twelve
calibration points can be entered.
If finished press ESC.
4.5.4 Cal. Interval
F2- Cal. Interval: calibration start time or interval between automatic calibrations:
You can enter either a time interval (1-24 hours) which determines the time between two
automatic calibrations or a fixed start time at which the automatic calibration will start daily.
The arrow indicates the active mode, it can be changed with the 68-keys.