UNI-T UT3000 series Instruções de Operação Página 67

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 98
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 66
6. Maintenance
page 67
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
6.4 Troubleshooting
6.4.1 Unusual Measuring Peaks
An example for a non-typical peak is shown in the picture below.
The peak is more of triangular shape and the signal drops to negative values after the peak.
Figure: Example for an non-typical desorption peak
The reason may be precipitation in the tubing at the outlet of the GoldTrap (= tubing which
connects GoldTrap and optical cell, see picture below).
In this case it is recommended to replace the contaminated tubing.
Figure: Tubing at outlet of GoldTrap. The arrow indicates the
position where precipitation might occur.
Vista de página 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 97 98

Comentários a estes Manuais

Sem comentários